Calibration method for electrode gains in an axially symmetric stripline BPM postprint

请选择邀稿期刊:
Abstract: The four electrodes in the stripline beam position monitor (BPM) for Hefei Light Source (HLS II) storage ring are of axially symmetric type. We have derived a new calibration method of electrode gains for this type stripline BPM. The gain fit error of different data grids was analyzed, and the � 5 mm by � 5 mm grid is the best. The electrode gains of two stripline BPMs (HLS II SR-BD-STLB1 and HLS II SR-BD-STLB2) were obtained based on offline calibrated data. The results show that data after fitting gains are improved, with the electrode gains being between 0.94 and 1.15.