A novel method for EPID transmission dose generation using Monte Carlo simulation and deep learning

请选择邀稿期刊:
Abstract: This study aimed to integrate Monte Carlo(MC) simulation with deep learning(DL)-based denoising techniques to achieve fast and accurate prediction of high-quality electronic portal imaging device (EPID) transmission dose (TD) for patient-specific quality assurance (PSQA). A total of 100 lung cases were used to obtain the noisy EPID TD by the ARCHER MC code under four kinds of particle numbers (1 × 106,1 × 107,1 × 108 and 1 × 109), and the original EPID TD was denoised by the SUNet neural network. The denoised EPID TD was assessed both qualitatively and quantitatively using the structural similarity (SSIM), peak signal-to-noise ratio (PSNR), and gamma passing rate (GPR) with respect to 1 × 109 as a reference. The computation times for both the MC simulation and DL-based denoising were recorded. As the number of particles increased, both the quality of the noisy EPID TD and computation time increased significantly (1 × 106: 1.12 s, 1 × 107: 1.72 s, 1 × 108: 8.62 s, and 1 × 109: 73.89 s). In contrast, the DL-based denoising time remained at 0.13–0.16 s. The denoised EPID TD shows a smoother visual appearance and profile curves, but differences between 1×106 and 1 × 109 still remain. SSIM improves from 0.61 to 0.95 for 1 × 106, 0.70 to 0.96 for 1 × 107, and 0.90 to 0.97 for 1 × 108. PSNR increases by > 20% for 1 × 106 and 1 × 107, and > 10% for 1 × 108. GPR improves from 48.47% to 89.10% for 1×106, 61.04% to 94.35% for 1×107, and 91.88% to 99.55% for 1×108. The method that combines MC simulation with DL-based denoising for EPID TD generation can accelerate TD prediction and maintain high accuracy, offering a promising solution for efficient PSQA.

Version History

[V1] 2025-12-31 21:53:44 ChinaXiv:202601.00010V1 Download
Download
Preview
Peer Review Status
Awaiting Review
License Information
metrics index
  •  Hits4341
  •  Downloads1418
Comment
Share