摘要: This paper reports an experimental investigation into the backscatter effects on gamma radiation monitors. The study addresses deviations in measurement readings observed in wall-mounted monitors compared to free-air devices due to backscatter phenomena. Reinforced concrete test blocks were employed to simulate operational environments, with measurements conducted using three detector types: a radiation protection level dosimeter with ionization chamber, an air-kerma radiation monitor with counter tube, and a radiation detector with NaI(Tl) scintillator. Analysis of backscatter factors at varying distances and orientations revealed that: (1) backscatter factors decreased with increasing source-to-surface distance, and (2) central irradiation generally yielded higher backscatter factors than non-central irradiation. The experimental data provide essential reference for formulating or revising pertinent technical standards.
-
来自:
梁, Dr. 杰
-
分类:
物理学
>>
核物理学
-
备注:
已向《Nuclear Science and Techniques》投稿
-
引用:
ChinaXiv:202507.00132
(或此版本
ChinaXiv:202507.00132V1)
DOI:10.12074/202507.00132
CSTR:32003.36.ChinaXiv.202507.00132
- 推荐引用方式:
梁, Dr. 杰,Wang, Mr. Ke-liang,Guo, Dr. Ji-ping,Zhang, Mrs. Fan,Wu, Ms. Ya-qin,Jun, Prof. Yang,Zhun, Prof. Ying-chun.The research of backscattering effect on the installed area monitors and/or warning devices for gamma radiation.中国科学院科技论文预发布平台.[DOI:10.12074/202507.00132]
(点此复制)